Physics > Instrumentation and Detectors
[Submitted on 30 Dec 2005]
Title:Phototube tests in the MiniBooNE experiment
View PDFAbstract: The MiniBooNE neutrino oscillation experiment at Fermilab uses 1520 8-inch PMTs: 1197 PMTs are Hamamatsu model R1408 and the rest are model R5912. All of the PMTs were tested to qualify for inclusion in the detector, sorted according to their charge and time resolutions and dark rates. Seven PMTs underwent additional low light level tests. The relative detection efficiency as a function of incident angle for seven additional PMTs was measured. Procedures and results are presented.
References & Citations
Bibliographic and Citation Tools
Bibliographic Explorer (What is the Explorer?)
Connected Papers (What is Connected Papers?)
Litmaps (What is Litmaps?)
scite Smart Citations (What are Smart Citations?)
Code, Data and Media Associated with this Article
alphaXiv (What is alphaXiv?)
CatalyzeX Code Finder for Papers (What is CatalyzeX?)
DagsHub (What is DagsHub?)
Gotit.pub (What is GotitPub?)
Hugging Face (What is Huggingface?)
Papers with Code (What is Papers with Code?)
ScienceCast (What is ScienceCast?)
Demos
Recommenders and Search Tools
Influence Flower (What are Influence Flowers?)
CORE Recommender (What is CORE?)
arXivLabs: experimental projects with community collaborators
arXivLabs is a framework that allows collaborators to develop and share new arXiv features directly on our website.
Both individuals and organizations that work with arXivLabs have embraced and accepted our values of openness, community, excellence, and user data privacy. arXiv is committed to these values and only works with partners that adhere to them.
Have an idea for a project that will add value for arXiv's community? Learn more about arXivLabs.