Physics > Medical Physics
[Submitted on 1 Aug 2013 (v1), last revised 12 Nov 2013 (this version, v2)]
Title:Monitor backscatter factors for the Varian 21EX and TrueBeam linear accelerators: measurements and Monte Carlo modeling
View PDFAbstract:Purpose/Objective: Backscattered radiation (BSR) into linac monitor chamber has to be accounted for in radiotherapy dose calculations. In Monte Carlo (MC) calculations, the BSR can be modeled explicitly, but only when treatment head geometry is available. In this study, monitor backscatter factors (MBSFs), defined as the ratio of the charge collected in the monitor chamber for a reference field to that of a given field, have been evaluated experimentally and incorporated into MC modelling of linacs with either known or unknown treatment head geometry.
Materials and methods: A telescopic technique similar to that by Kubo (1989) was used. However, instead of lead slits, a 1.8 mm diameter collimator and a small (2 mm diameter) detector positioned at extended SDD were used. This setup provided a field of view to the source of less than 3.1 mm and allowed for MBSF measurements of open fields from 1 x 1 to 40 x 40 cm2. MBSFs were also explicitly modeled in MC calculations using BEAMnrc and DOSXYZnrc codes for 6MV and 18MV beams of a Varian 21EX linac. A method for deriving the Dchforward values that are used in MC absolute dose calculations was demonstrated. These values were derived from measured MBSFs for two 21EX and four TrueBeam energies.
Results: MBSFs were measured for 6MV and 18MV beams from Varian 21EX, and for 6MV, 10MV-FFF, 10MV, and 15MV beams from Varian TrueBeam linacs. For the open field sizes modeled in this study for the 21EX, the measured MBSFs agreed with MC calculated values within combined statistical (0.4%) and experimental (0.2%) uncertainties. Variation of MBSFs across field sizes was about a factor of two smaller for the TrueBeam compared to 21EX Varian linacs.
Conclusions: Measured MBSFs and the derived Dchforward factors allow for the incorporation of the BSR effect into accurate radiotherapy dose calculations without explicit backscatter modelling.
Submission history
From: Sergei Zavgorodni [view email][v1] Thu, 1 Aug 2013 22:12:52 UTC (1,182 KB)
[v2] Tue, 12 Nov 2013 20:43:24 UTC (786 KB)
Current browse context:
physics.med-ph
Change to browse by:
References & Citations
Bibliographic and Citation Tools
Bibliographic Explorer (What is the Explorer?)
Connected Papers (What is Connected Papers?)
Litmaps (What is Litmaps?)
scite Smart Citations (What are Smart Citations?)
Code, Data and Media Associated with this Article
alphaXiv (What is alphaXiv?)
CatalyzeX Code Finder for Papers (What is CatalyzeX?)
DagsHub (What is DagsHub?)
Gotit.pub (What is GotitPub?)
Hugging Face (What is Huggingface?)
Papers with Code (What is Papers with Code?)
ScienceCast (What is ScienceCast?)
Demos
Recommenders and Search Tools
Influence Flower (What are Influence Flowers?)
CORE Recommender (What is CORE?)
arXivLabs: experimental projects with community collaborators
arXivLabs is a framework that allows collaborators to develop and share new arXiv features directly on our website.
Both individuals and organizations that work with arXivLabs have embraced and accepted our values of openness, community, excellence, and user data privacy. arXiv is committed to these values and only works with partners that adhere to them.
Have an idea for a project that will add value for arXiv's community? Learn more about arXivLabs.