Physics > Medical Physics
[Submitted on 1 Dec 2021]
Title:Open-bore vertical MRI scanners generate significantly less RF heating around deep brain stimulation leads compared to horizontal scanners
View PDFAbstract:Objectives Studies that assess magnetic resonance imaging (MRI) induced radiofrequency (RF) heating of the tissue in the presence of an active electronic implant are mostly performed in horizontal, closed-bore scanners. Vertical, open-bore MRI systems have a 90° rotated magnet and generate a fundamentally different RF field distribution in the body, yet little is known about the RF heating of deep brain stimulation (DBS) systems in this class of scanners. Here, we investigated whether RF heating of DBS devices was significantly different in a vertical, open-bore MRI scanner compared to a horizontal, closed-bore MRI scanner. Materials and Methods In this phantom study, RF heating around the lead of a commercial DBS system implanted in an anthropomorphic phantom was evaluated in a 1.2 T vertical open-bore scanner (Oasis, Fujifilm Healthcare) and a 1.5 T horizontal closed-bore scanner (Aera, Siemens Healthineers). DBS devices were implanted following 30 realistic lead trajectories. Electromagnetic simulations were performed to assess the specific absorption rate (SAR) of RF energy around leads with different internal structures. Results When controlling for B1+rms, temperature increase around the lead-tip was significantly lower at the vertical scanner compared to the horizontal scanner (p-value=9.1x10-7). Electromagnetic simulations demonstrated up to a 14-fold reduction in the maximum 0.1g-averaged SAR deposited in the tissue surrounding the lead-tip in a vertical scanner compared to a horizontal scanner for leads with straight and helical internal wires. Conclusions RF experiments and electromagnetic simulations demonstrated consistently lower RF heating and power deposition around the DBS lead-tip at the vertical scanner compared to the horizontal scanner. Our simulation results suggest that this trend in heating may potentially extend to leads from other manufacturers.
Current browse context:
physics
Change to browse by:
References & Citations
Bibliographic and Citation Tools
Bibliographic Explorer (What is the Explorer?)
Litmaps (What is Litmaps?)
scite Smart Citations (What are Smart Citations?)
Code, Data and Media Associated with this Article
CatalyzeX Code Finder for Papers (What is CatalyzeX?)
DagsHub (What is DagsHub?)
Gotit.pub (What is GotitPub?)
Papers with Code (What is Papers with Code?)
ScienceCast (What is ScienceCast?)
Demos
Recommenders and Search Tools
Influence Flower (What are Influence Flowers?)
Connected Papers (What is Connected Papers?)
CORE Recommender (What is CORE?)
arXivLabs: experimental projects with community collaborators
arXivLabs is a framework that allows collaborators to develop and share new arXiv features directly on our website.
Both individuals and organizations that work with arXivLabs have embraced and accepted our values of openness, community, excellence, and user data privacy. arXiv is committed to these values and only works with partners that adhere to them.
Have an idea for a project that will add value for arXiv's community? Learn more about arXivLabs.